Envea

Home > Product > AF22e

AF22e

e-Series Sulfur dioxides Analyzer SO₂

UV Fluorescence gas monitor offering excellent performances for the monitoring of SO2 in the range 0-10 ppm or 0-1 ppm. Option: H2S / TRS monitoring.

 

Eco-designed, reference air quality monitor, QAL1 certified by the TÜV, US-EPA approved. Utilizes the standard method for the measurement of sulfur dioxides (EN 14212). With its ultra low power consumption & 24 V supply (option), the AF22e is the sole SO2 monitor on the market able to measure without being connected to the power grid.

FEATURES AND BENEFITS

  • Superior metrological performances for SO2 measurements in ranges 0-10 ppm or 0-1 ppm
  • Innovative design of the optical module for excellent sensitivity and signal stability
  • Real-time calibration graph, animated synoptic, auto-diagnostic, control and maintenance data screens can be displayed while the instrument is operating
  • Embedded Communication Protocol for XR® Software with automatic recognition and configuration
  • Able to withstand temperatures up to 45-50 °C without air conditioning
  • Special architecture means that no additional pump is necessary in case of permeation bench (in option)
  • Automatic recognition of plugged electronic boards or optional devices: plug & play principle
  • Smart, connected instrument with embedded web server for the easiest, full remote control and operation of the analyzer from any PC, tablet or smartphone (configuration, calibration, test, diagnostic parameters…)
  • Integrated web-server with full remote emulation of the analyzer through the ENVEA Connect free app

 

MAIN OPTIONS:

  • Analyzer without screen for reduced environmental impact
  • Additional parameters: SO2, H2S, CH2SH, COS, CS2, (CH2)2S…
  • 24 V Power supply PLUS extended temperature range 0-50°C
  • Built-in permeation bench with SO2 tube (additional pump not required)
  • Internal converter for H2S monitoring (range 0-1000 ppb)
  • CTRS External module, to be used in combination with the AF22e SO2 analyzer for the continuous measurement of TRS compounds or cyclic measurement SO2/TRS
 

MAIN APPLICATIONS

  • Continuous indoor and outdoor air quality monitoring
  • Stationary and mobile AQMS laboratories
  • Industrial fence-line monitoring
  • Continuous emissions monitoring (CEM) by dilution
  • Laboratory & field studies on SO/ H2S / TRS effects…
  • Background (urban or suburban) Rural, Traffic, Kerbside measurement campaigns and monitoring studies
 
 

TECHNICAL INFORMATION

Measured parameter

SO2

Technology

UV Fluorescence

US-EPA Approval / QAL 1 Certification

YES

Measurement ranges

0-1 ppm or 0-10 ppm user-defined

Measurement units

ppb, ppm, µg/m3 user-defined

Lower Detectable Limit (2σ)

< 0.4 ppb

Noise (σ)

< 0.2 ppb

Zero drift

< 1 ppb / 24h

Span drift

< 0.5% / 7 days

Linearity

± 1%

Sample flow rate

20L/h

Sample pump

internal

Temperature range

0°C to +40°C

Data storage

1 year

Included I/O

1LAN/3USB/WIFI/2DRY

Weight (Kg)

9,8

Dimensions mm (LxWxH)

483 x 545 x 133

Chassis

19″ Rack, 3U

Power supply

115 V, 60 Hz – 230 V, 50 Hz
24 V optional

Pressure and temperature compensation

YES

APPROVALS & CERTIFICATIONS

COMPLIANCE WITH:

  • 2008/50/EC, EN 14212, EN 15267, 40 CFR PART 53 SUB B and SUB C
  • QAL 1 certified by the TÜV to N° 0000051690
  • U.S. EPA approved as compliant with EQSA-0802-149

ENVEA monitoring solutions are European QAL 1 certified and in compliance with the latest international regulations & standards.
They are also approved and certified by various laboratories and organizations around the world such as: US EPA, TÜV, GOST, CEN, JQA, CNEMC, JMOE, KTL, CNSA, LCSQA…