Envea

Home > Product > PA-2 / PA2-Gold

PA-2 / PA2-Gold

MERCURY ANALYZER FOR WATER AND AQUEOUS SOLUTIONS IN PROCESS APPLICATIONS

The mercury analyzer PA-2 is used for continuous monitoring of mercury concentrations in industrial processes. The PA-2 was specially designed for process analysis and operation under harsh industrial condition. The system comes with either the VM-3000 or the UT-3000 photometer to meet different requirements regarding accuracy and sensitivity of measurements.

FEATURES AND BENEFITS

  • Fully automatic system
  • Easy menu-driven operation
  • Proven and reliable detection method: AAS
  • Measuring ranges from 10 ppt to 100 ppm
  • High flexibility of sample pretreatment
  • Suited for complex sample compositions
  • Low reagent consumption
  • Corrosion-protected construction
  • Automatic self diagnosis system for reliable operation

MAIN APPLICATIONS

  • Quality control in chlorine-alkali plants
  • Monitoring of scrubber water of waste incinerators and power plants
  • Control of industrial sewage
  • Control of purification plants
  • Quality control of sulphuric acid
  • Quality control of drinking water

TECHNICAL INFORMATION

Measuring principle

Amalgamation on MI GoldTrap (PA-2 Gold only), cold vapor atomic absorption (CVAAS) at 253.7 nm

UV-Source

Electrodeless low pressure mercury lamp (EDL)

Stabilization method

Reference beam method

Optical cell

Fused silica (Suprasil), lapprox. 230mm, heated approx. 55°

Measuring ranges PA-2

0.1 … 1 µg/l to 0…10 mg/l

Measuring ranges PA-2 Gold

0.01 … 1 μg/l to 0.1 … 10 μg/l

Sensitivity

> 0.01 µg/l

Response time

approx. 1 minute

Sample flow

approx. 20 l/h

Carrier gas

compressed air, filtered, approx. 30l/h, 1bar

Reduction reagent

Tin-II-chloride or Sodiumhydroborate

Reagent consumption

approx. 3l/week for each reagent

Sample digestion

HCl or H2SO4, KMnO4 or H2O2 or Fenton’s Reagent or NaClO3; depending on sample composition

Signal outputs

analogue: 4 – 20 mA (500 Ω max.);
serial: RS232 / USB/td >

Status output

Calibration; Autozero; Service; Malfunction; according to NAMUR 64 recommendations

Housing

Fibreglass-reinforced polyester, for use in highly corrosive environment – Reagent storage cabinet made of chemically resistent plastic material/td >

Protection class

IP 66 (EN 60529/ NEMA 3; 3R; 4; 4X; 12; 13)

Operating temperature

0 to 40°C

Sample temperature

65°C

Power supply

230 VAC / 50 Hz (110 VAC / 60 Hz option)

El. Power consumption

750 VA max. (PA-2 Gold, heating peak)

Dimensions (W x H x D)

62 x 78 x 33 cm

Weight

approx. 50 kg

APPROVALS & CERTIFICATIONS

Certifiet:

  • Our products are manufactured according to the ISO 9001 quality regulations.
  • CE approved

CUSTOMER STORIES