Envea

Home > Product > UT-3000

UT-3000

TOTAL GASEOUS MERCURY (TGM) ULTRATRACER

Measuring of Total Gaseous Mercury (TGM) in air and other gases. Reliable for measurements at ultra-trace levels (sub-ng/m3) according to EN 15852.

 

The UT-3000 operates automatically. All functions are controlled by an embedded microprocessor. When measurements are carried out, all parameters and results are stored in the integrated data logger. 

 

With standard settings the data logger has a recording capacity of 5000 measurements (approx. 2 months).

FEATURES AND BENEFITS

  • Offers sensitivity and detection limits far superior to other mercury analyzers on the market using similar technologies
  • Unlike with the fluorescence detection method, the analyzer does not require expensive carrier gases
  • Insensitive to interferences caused by the quenching effect
  • Automatic Operation
  • Automatic sample flow control
  • AutoRange function
  • Manual or automatic calibration modules (optional accessories)
  • Integrated data logger for more than 5000 measurements
  • Multiplexer operation (option)
  • Special version UT-3000 NG mobile(option) – all in one system for mercury measurements in natural gas
 
 

MAIN APPLICATIONS

  • Continuous outdoor and indoor air quality monitoring
  • Stationary and mobile AQMS laboratories
  • Industrial fence-line and field monitoring
  • Environmental pollution source tracking
  • Studies of mercury dispersion in the atmosphere
  • Natural gas and derivatives
  • Emission monitoring
  • Process monitoring
  • Plume profiling
  • Studies of interaction between atmosphere and surfaces
  • Soil and ground degassing studies
 
 

TECHNICAL INFORMATION

Measuring principle

Amalgamation (MI GoldTrap), cold vapor atomic absorption (CVAAS) at 253.7 nm

Measured component

Total Gaseous Mercury (TGM)

UV-Source

Electrodeless low pressure mercury lamp (EDL)

Stabilization method

Reference beam method

Optical cell

Fused silica (Suprasil), l: approx. 230mm, heated approx. 45°

Detection limit

0,1 ng/m3 corresponding to 0,5 pg Hg absolutely

Sample volume

0,1 l to 10 l

Sampling duration

9 sec … 15 min

Measurement cycle time

3 … 99 min

Measuring ranges

at 10 l sample volume: 0,1 ng/m3 to 1000 ng/m3; at 1 l sample volume: 1 ng/m3 to 10 000 ng/m3

Data logger function

integrated, for up to 5 000 measurements

Calibration

manually – by injecting calibration gas into the calibration port (syringe optionally available); automatic -with calibration gas unit (option)

Signal outputs

analogue: 4 – 20 mA (500 Ω max.); serial: RS232 / USB

Status output

3 x pairs of relay contacts (dry contacts)

Ambient temperature range allowed

0 to 40°C

Power supply

230 VAC / 50 Hz -110 VAC / 60 Hz

El. Power consumption

250 VA max. (heating peak)

Battery power

supply pack optionally available

Dimensions (W x H x D)

45 x 15 x 35 cm (without filter)

Weight

approx. 9 kg

APPROVALS & CERTIFICATIONS

Certified:

  • Compliance with DIN EN 15852 – Ambient air quality – Standard method for the determination of total gaseous mercury
  • Our products are manufactured according to the ISO 9001 quality regulations.
  • CE approved