Measuring of Total Gaseous Mercury (TGM) in air and other gases. Reliable for measurements at ultra-trace levels (sub-ng/m3) according to EN 15852.
The UT-3000 operates automatically. All functions are controlled by an embedded microprocessor. When measurements are carried out, all parameters and results are stored in the integrated data logger.
With standard settings the data logger has a recording capacity of 5000 measurements (approx. 2 months).
Amalgamation (MI GoldTrap), cold vapor atomic absorption (CVAAS) at 253.7 nm
Total Gaseous Mercury (TGM)
Electrodeless low pressure mercury lamp (EDL)
Reference beam method
Fused silica (Suprasil), l: approx. 230mm, heated approx. 45°
0,1 ng/m3 corresponding to 0,5 pg Hg absolutely
0,1 l to 10 l
9 sec … 15 min
3 … 99 min
at 10 l sample volume: 0,1 ng/m3 to 1000 ng/m3; at 1 l sample volume: 1 ng/m3 to 10 000 ng/m3
integrated, for up to 5 000 measurements
manually – by injecting calibration gas into the calibration port (syringe optionally available); automatic -with calibration gas unit (option)
analogue: 4 – 20 mA (500 Ω max.); serial: RS232 / USB
3 x pairs of relay contacts (dry contacts)
0 to 40°C
230 VAC / 50 Hz -110 VAC / 60 Hz
250 VA max. (heating peak)
supply pack optionally available
45 x 15 x 35 cm (without filter)
approx. 9 kg