Measuring principle
Amalgamation (MI GoldTrap), cold vapor atomic absorption (CVAAS) at 253.7 nm
UV-Source
Electrodeless low pressure mercury lamp (EDL)
Stabilization method
Optical cell
Fused silica (Suprasil), l: approx. 230mm, heated approx. 45°
Measured component:
Total Gaseous Mercury (TGM)
Measuring ranges
Lowest range: 0.1 ng/m3 – 1000 ng/m3 (at 10 l sample gas volume); medium range: 1 ng/m3 – 10 000 ng/m3 (at 1 l sample gas volume); screening mode: 10 ng/m3 – 30 000 ng/m3 (at 0.1 l sample gas volume)
Detection limit
1 pg absolutely or 0.1 ng/m3
Linearity over the 0.2 to 15000 ng/m3 range
Reproducibility
Signal outputs
Status output
3 x pairs of relay contacts (dry contacts)
Max. sample temperature
Operating temperature
Power supply
230 VAC / 50 Hz (110 VAC / 60 Hz option)
Power consumption
System: 425 VA; (UT-3000: 250 VA max); (NG Sampling System 3000: 175 VA)
Dimensions (W x H x D)
Weight