Autonomous networks of sensor-based mini-stations
Cairnet® device is a real-time standalone and networkable air-monitoring station including up to six Cairsens® microsensors. It is powered by solar panels and offers cellular communication.
Thanks to its integrated manifold allowing a dynamic sampling, the measures are accurate and reliable. Electronic compounds are sealed & protected from moisture and corrosive ambient air.
Compact, easily deployable and autonomous thanks to its solar panels and battery, Cairnet® enables you to cost-effectively monitor dust and gas and gives a complete picture of the environmental impact of your operations.
It offers unprecedented flexibility in producing accurate and dynamic air quality measurements across a broad range of industries and applications.
Real time continuous measurement and monitoring of up to 10 gaseous and particulate pollutants simultaneously, among: H2S/CH4S, NH3, nmVOC, O3+NO2,NO2, CO, SO2, PM10, PM2.5 & PM1
- Measurement of environmental parameters: temperature, relative humidity and pressure
- Plug & Play Network: automatic pairing (Cairsens® / Cairnet® / Caircloud®)
- Dynamic sampling and improved protection against moist and/or corrosive environments
- Optional: Ultrasonic anemometer to measure wind direction and speed. Autonomous
thanks to its integrated solar panel and battery
- Automated data saving (micro SD card) and push : no consequence in case of lost
communication service - Communication frequency and measured data volume adapted automatically to the
autonomy of the station - Very high sensitivity to capture low level gas and particulate concentrations
- Modular, easy to use and move on-site: no cables
- Cairnet® requires only annual maintenance: when sensors should be renewed
- Operating cost savings: process adjustment & Improvement of local communication
(neighbors & authorities) - Possibility to set up hybrid AQMS networks (reference stations & mini-stations)
- Odor monitoring: WWTP, recycling, pulp and paper manufacturing, sewerage treatment facilities, refineries
- Indoor and outdoor air quality monitoring: smart cities, road-side & tunnels, schools, airports, ship terminals…
- Process leak detection and monitoring of fugitive emissions: quarries, storage facilities, mines, manufacturing plants
- Forecasting of industrial fence line emissions
- Environmental impact assessments
- Health and safety: mines, industrial sites, construction
- Mapping and modelling pollution sources
- Every Cairnet measurement point consists of:
- Solar panel and its battery
- Wireless communication module
- Weather protection enclosure including up to 6 Cairsens miniature sensors, each measuring one or two pollutants. The sensors are to be selected from the list, function of your specific requirements.
- Optional: ultrasonic anemometer. If this option is selected, it replaces a Cairsens® sensor
Measurable parameters | ||||
---|---|---|---|---|
Pollutant | Range (ppb) |
Certified detection limit (ppb)* |
Resolution (ppb) |
Order code |
NO2 | 0-250 | 20 | 1 | A40-0405 |
O3 / NO2 | 0-250 | 20 | 1 | A40-0406 |
SO2 | 0-1,000 | 50 | 1 | A40-0407 |
CO | 0-20,000 | 50 | 1 | A40-0404 |
H2 S / CH4S |
0-1,000 0-20,000 0-200,000 |
10 30 200 |
1 | A40-0401 A40-0402 A40-0403 |
NH3 | 0-25,000 | 500 | 1 | A40-0408 |
nmVOC | 0-16,000 0-2,000 |
500 200 |
1 | A40-0409 A40-0410 |
PM10/PM2.5/PM1 | 0-1,000 µg/m3 | <5 µg/m3 | 0.01 µg/m3 | A40-0414 |
Technical specifications | |||||
---|---|---|---|---|---|
Power supply | 8 to 30 V DC / 2.5 A or battery (included) | ||||
Battery included | 3.7 V – 22 Ah, Li-Ion rechargeable on 18VDC / 2A or via solar pannels (option) | ||||
Solar Panels Kit (option) | 27 Watts. Mounting bracket included | ||||
Control & Data Treatment Board | Internal microprocessor for data acquisition, power and communication managment etc. Embedded Real-Time-Clock (auto-adjusted at every communication) |
||||
Wireless communication | Cellular technology LTE – 3G / 4G or more (SIM card optional for European countries, not provided for other countries) Regulatory compliance: R&TTE directive 1999/5/EC, Japan JRF/JPA – FCC – IC |
||||
Data storage | • Internal storage on micro SD card of all data in case of loss of cellular communication • Cairsens® internal memory |
||||
Mounting | Fixation kit for pole (⌀50 mm max ) included. | ||||
Dimensions of the Cairnet® housing with fixation kit & antenna | 300 x 215 x 257 mm (L x H x W) | ||||
Dimensions of the solar Panels with its fixation kit | 800 x 410 x 100 mm (L x H x W) | ||||
Weight of the Cairnet® housing | 4 Kg | ||||
Weight of the Solar Panels kit | 4.9 Kg | ||||
Environmental using conditions | -20°C to +50°C / RH 10% to 90% | ||||
Option: ultrasonic anemometer (article code A40-0220) |
|
- Technology and frequency bands: LTE (1800, 2100, 2600 MHz), WCDMA (800,1700, 2100 MHz), GPRS/Edge (850 MHz, 900 MHz, 1800 MHz, 1900 MHz)
- Regulatory compliance: R&TTE directive 1999/5/EC Japan JRF/JPA – FCC – IC