Sensor-based mini-stations
Cairnet® DATA+ is a real-time air quality monitoring station including up to six Cairsens® micro-sensors.
It also monitors environmental parameters such as wind speed and direction, T°C, RH% and Pa. The device has been designed to leverage local communication and features main power supply as well as cellular communication. It can acquire data from complementary sources such as “all in one” weather stations, noise sensors, etc.
Cairnet® DATA+ can be easily connected to our cloud solutions or customer software plateform by using APIs.
Real time continuous measurement and monitoring of up to 10 gaseous and particulate pollutants simultaneously, among: H2S/CH4S, NH3, nmVOC, O3+NO2, NO2, CO, SO2, PM10, PM2.5 & PM1
- Local and remote communication: 4G via local server, Wi-Fi, TCP-IP Modbus, 4-20 mA
- Indoor/outdoor compact & easily deployable housing
- Designed to acquire data from complementary sources such as “all in one” weather stations, noise sensors, etc.
- Integrated manifold: dynamic sampling
- Limited maintenance: annual replacement of sensor only
- Protection of electronic components against the effects of corrosion
- Data security and privacy thanks to direct and local communication to the reporting platform
- Accurate and reliable measurements across a broad range of industries and applications
- Operating cost savings: process adjustment & improvement of local communication
- Gives a complete picture of the environmental impact of your operations
- WWTP, recycling
- Pulp and paper manufacturing
- Refineries
- Manufacturing plants and gas transport
- Urban areas/cities: road traffic, schools, tunnels
- Complement to Reference Air Quality Monitoring Network
- Industrial areas: airports, ship terminals, manufacturing plants
Measurable parameters | ||||
---|---|---|---|---|
Pollutant | Range (ppb) |
Certified detection limit (ppb)* |
Resolution (ppb) |
Order code |
NO2 | 0-250 | 20 | 1 | A40-0405 |
O3 / NO2 | 0-250 | 20 | 1 | A40-0406 |
SO2 | 0-1,000 | 50 | 1 | A40-0407 |
CO | 0-20,000 | 50 | 1 | A40-0404 |
H2 S / CH4S |
0-1,000 0-20,000 0-200,000 |
10 30 200 |
1 | A40-0401 A40-0402 A40-0403 |
NH3 | 0-25,000 | 500 | 1 | A40-0408 |
nmVOC | 0-16,000 0-2,000 |
500 200 |
1 | A40-0409 A40-0410 |
PM10/PM2.5/PM1 | 0-1,000 µg/m3 | <5 µg/m3 | 0.01 µg/m3 | A40-0414 |
Technical specifications | |||||
---|---|---|---|---|---|
Power supply (included) | 12 VDC / 2.5 A | ||||
Internal data logger | Microprocessor for data acquisition, power and communication management, etc. Based on eSAM software |
||||
Wireless communication | Wi-Fi, Cellular technology LTE / 4G or more (SIM card, not provided) Regulatory compliance: R&TTE directive 1999/5/EC, Japan JRF/JPA – FCC – IC |
||||
Data storage |
|
||||
Mounting | Fixation kit for pole (⌀ 50 mm max ) included | ||||
Dimensions of the Cairnet® housing with fixation kit & antenna | 300 x 215 x 257 mm (L x H x W) | ||||
Dimensions of the outdoor power supply (included) | 150 x 50 x 150 mm (L x H x W) | ||||
Weight of the Cairnet® housing | 4 Kg | ||||
Weight of the outdoor power supply | 1 Kg | ||||
Environmental operating conditions | -20°C to +50°C / RH 10% to 90% | ||||
Option: ultrasonic anemometer (article code A40-0220) |
|
||||