Total Gaseous Mercury (TGM) ultratracer

UT-3000

Measuring of Total Gaseous Mercury (TGM) in air and other gases. Reliable for measurements at ultra-trace levels (sub-ng/m3) according to EN 15852.

The UT-3000 operates automatically. All functions are controlled by an embedded microprocessor. When measurements are carried out, all parameters and results are stored in the integrated data logger. 

With standard settings the data logger has a recording capacity of 5000 measurements (approx. 2 months).

  • Offers sensitivity and detection limits far superior to other mercury analyzers on the market using similar technologies
  • Unlike with the fluorescence detection method, the analyzer does not require expensive carrier gases
  • Insensitive to interferences caused by the quenching effect
  • Automatic Operation
  • Automatic sample flow control
  • AutoRange function
  • Manual or automatic calibration modules (optional accessories)
  • Integrated data logger for more than 5000 measurements
  • Multiplexer operation (option)
  • Special version UT-3000 NG mobile (option) – all in one system for mercury measurements in natural gas
  • Continuous outdoor and indoor air quality monitoring
  • Stationary and mobile AQMS laboratories
  • Industrial fence-line and field monitoring
  • Environmental pollution source tracking
  • Studies of mercury dispersion in the atmosphere
  • Natural gas and derivatives
  • Emission monitoring
  • Process monitoring
  • Plume profiling
  • Studies of interaction between atmosphere and surfaces
  • Soil and ground degassing studies
Technical Specifications
Measuring principle Amalgamation (MI GoldTrap), cold vapor atomic absorption (CVAAS) at 253.7 nm
Measured component Total Gaseous Mercury (TGM)
UV-Source Electrodeless low pressure mercury lamp (EDL)
Stabilization method Reference beam method
Optical cell Fused silica (Suprasil), l: approx. 230mm, heated approx. 45°
Detection limit 0,1 ng/m3 corresponding to 0,5 pg Hg absolutely
Sample volume 0,1 l to 10 l
Sampling duration 9 sec … 15 min
Measurement cycle time 3 … 99 min
Measuring ranges at 10 l sample volume: 0,1 ng/m3 to 1000 ng/m3; at 1 l sample volume: 1 ng/m3 to 10 000 ng/m3
Data logger function integrated, for up to 5 000 measurements
Calibration manually – by injecting calibration gas into the calibration port (syringe optionally available); automatic -with calibration gas unit (option)
Technical data
Signal outputs analogue: 4 – 20 mA (500 Ω max.); serial: RS232 / USB
Status output 3 x pairs of relay contacts (dry contacts)
Ambient temperature range allowed 0 to 40°C
Power supply 230 VAC / 50 Hz -110 VAC / 60 Hz
El. Power consumption 250 VA max. (heating peak)
Battery power supply pack optionally available
Dimensions (W x H x D) 45 x 15 x 35 cm (without filter)
Weight approx. 9 kg
  • Compliance with DIN EN 15852 – Ambient air quality – Standard method for the determination of total gaseous mercury
  • Our products are manufactured according to the ISO 9001 quality regulations.
  • CE approved