e-Series Nitrogen Oxides Analyzer NO, NOx & NO₂
Precise PM1, PM2.5, PM4.25, PM10 and TSP monitoring in indoor ambient air.
Eco-designed, reference air quality monitor, based on the chemiluminescence technology, the standard method for the measurement of nitrogen oxides (EN 14211). QAL1 certified by the TÜV, US-EPA approved. With its ultra low power consumption & 24 V supply (option), the AC32e is the sole NOx analyzer on the market able to measure without being connected to the power grid.
Multi-screen remote access via Wifi or Ethernet using the ENVEA Connect™ application for control, diagnostics, software upgrade
- Features & Benefits
- Main applications
- Technical Specification
- Technical Data
- Approvals & Certifications
- Downloads
- Excellent sensitivity & metrological performances for NO, NO2 and NOx measurements in the range 0-1 ppm and optionally NH3
- Real-time calibration graph, animated synoptic, auto-diagnostic, control and maintenance data screens can be displayed while the instrument is operating
- Service assistance inside: detects early signs of trouble, allows predictive maintenance, identifies the needed service and guides service operations step by step: increased productivity on site, reduced downtime, more efficiency, less training…
- Embedded Communication Protocol for XR® Software with automatic recognition & configuration
- Able to withstand temperatures up to 45-50 °C without air conditioning
- Smart, connected instrument with embedded web server for the easiest, full remote control and operation of the analyzer from any PC, tablet or smartphone (configuration, calibration, test, diagnostic parameters…)
- Continuous indoor and outdoor air quality monitoring
- Stationary and mobile AQMS laboratories
- Ammonia impact assessments from intensive farming & livestock installations
- Industrial fence-line monitoring
- Continuous emissions monitoring (CEM) by dilution
- Background (urban or suburban), Rural, Traffic, Kerbside measurement campaigns and monitoring studies
- Laboratory and field studies on pollution effects
Technical Specifications | |
---|---|
Measured parameter | NO, NO₂ & NOx (NH3 optional) |
Technology | Chemiluminescence |
US-EPA Approval / QAL 1 Certification | YES |
Measurement ranges | 0-1 or 0-10 ppm for NO 0-1 ppm for NO2 ; 0-10 ppm for NO2 upon request. |
Measurement units | ppb, ppm, µg/m3 (user-selectable) |
Lower Detectable Limit (2σ) | < 0.2 ppb |
Noise (σ) | 0.1 ppb |
Zero drift | < 1 ppb / 24h |
Span drift | < 1% / 7 days |
Linearity | 1% |
Sample flow rate | 0.66 l/min (1 l/min with dryer) (with CHN3 option, flow rate limited to 0.33 l/min) |
Sampling pump | External |
Temperature range | 0°C to +40°C / optional 0°C to +50°C |
Data storage | 1 year |
Included I/O | 1LAN/3USB/WIFI |
Weight (Kg) | 10.3 (+ 4,6 pompe) |
Dimensions mm (LxWxH) | 483 x 606 x 133 |
Chassis | 19″ Rack, 3U |
Power supply | ‘100~250Vac, 50/60Hz + ground / 24 V optional |
Pressure and temperature compensation | YES |
- MAIN OPTIONS:
- Special version with extended temperature range and 24 V power supply for use in mobile or fixed AQMS solar powered or using green energy
- 7’’ TFT color touch screen
- WiFi module (in standard with the no-screen version)
- RS232 or RS485 serial interface (via USB port)
- External module for NH3 measurement (0-1000 ppb)
- Built-in permeation bench with NO2 tube
- Sample dryer
- External opto-isolated I/O interface with:
4 independent analog inputs
4 independent analog outputs
4 remote control inputs
6 dry contacts outputs
They are also approved and certified by various laboratories and organizations around the world such as: US EPA, TÜV, GOST, CEN, JQA, CNEMC, JMOE, KTL, CNSA, LCSQA…
- COMPLIANCE WITH:
- ISO 7996, EN 14211 (2012), EN 15267 (2009), 40 cfr part 53 & part 58
- QAL 1 certified by the TÜV to N° 0000053805
- U.S. EPA approved as compliant with RFNA-0118-249