Measuring principle
Amalgamation on MI GoldTrap (PA-2 Gold only), cold vapor atomic absorption (CVAAS) at 253.7 nm
UV-Source
Electrodeless low pressure mercury lamp (EDL)
Stabilization method
Optical cell
Fused silica (Suprasil), lapprox. 230mm, heated approx. 55°
Measuring ranges PA-2
0.1 … 1 µg/l to 0…10 mg/l
Measuring ranges PA-2 Gold
0.01 … 1 μg/l to 0.1 … 10 μg/l
Sensitivity
Response time
Sample flow
Carrier gas
compressed air, filtered, approx. 30l/h, 1bar
Reduction reagent
Tin-II-chloride or Sodiumhydroborate
Reagent consumption
approx. 3l/week for each reagent
Sample digestion
HCl or H2SO4, KMnO4 or H2O2 or Fenton’s Reagent or NaClO3; depending on sample composition
Signal outputs
analogue: 4 – 20 mA (500 Ω max.);
serial: RS232 / USB/td >
Status output
Calibration; Autozero; Service; Malfunction; according to NAMUR 64 recommendations
Housing
Fibreglass-reinforced polyester, for use in highly corrosive environment – Reagent storage cabinet made of chemically resistent plastic material/td >
Protection class
IP 66 (EN 60529/ NEMA 3; 3R; 4; 4X; 12; 13)
Operating temperature
Sample temperature
Power supply
230 VAC / 50 Hz (110 VAC / 60 Hz option)
El. Power consumption
750 VA max. (PA-2 Gold, heating peak)
Dimensions (W x H x D)
Weight