QAL 1 certified for continuous, accurate and reliable measurement of total mercury emissions, even in very low concentrations and within complex matrices (SO2, NOx, HCl, etc.).
With over 20 years experience, Mercury Instruments (part of ENVEA Group) were the first to develop mercury monitors and mercury analyzers using a dry thermocatalytic method instead of wet chemistry.
The SM-4 Mercury Monitor is QAL1 certified and is ready to meet future regulatory requirements (Minamata convention & European directives for thermal power plants and incinerators).
Main options:
thermocatalytic reduction at stack, Amalgamation (MI GoldTrap), cold vapor atomic absorption (CVAAS) at 253.7 nm
Total mercury (ionic and elemental speciation as option)
Electrodeless low pressure mercury lamp (EDL)
Reference beam method
Fused silica (Suprasil), l: approx. 230mm, heated approx. 45°
0.05 μg/m3 – 500 μg/m3 Hg (up to 1000 µg/m3 as an option)
0.0001 μg/m3 (Detector); < 0.01 μg/m3 (System)
180 – 360 sec
– automatic, with built-in calibration gas generator for elementary Hg (option); – automatic, with built-in calibration gas generator for ionic Hg (option); – manual, with built-in heated fitting to connect externally provided calibration gas e.g.HOVACAL
analogue: 4-20 mA (500 Ω max.); serial: RS232; Modbus RTU/RS485 (option); Ethernet (option)
3 x pairs of relay contacts (dry contacts)
Modem and SM-4 software (option)
250 °C (others upon request)
-5 to 40 °C
up to 30 m (up to 100 m option)
230 VAC / 50 Hz (110 VAC / 60 Hz option)
Analyzer cabinet: 450 VA; Probe-converter unit: 1000 VA; heated sample line: 30 VA / meter
Analyzer cabinet: 161 x 80 x 60 cm; Probe converter unit: 36 x 36 x 66 cm; Probe controller: 48 x 44 x 27 cm
Analyzer cabinet (including automatic calibrator): 130 kg; Probe converter unit: 28 kg; Probe controller: 14 kg
Approved and QAL 1 certified by TÜV Rheinland Energie und Umwelt GmbH, Köln, (QAL1)
Our mercury analyzers are manufactured according to the ISO 9001 quality regulations.
CE approved