ENVEA's LabAnalyzer 254 mercury monitoring system, designed for precise and reliable laboratory measurements of mercury concentrations. The compact and advanced design features intuitive controls and high-performance capabilities, ideal for environmental and industrial applications requiring accurate mercury analysis.

LabAnalyzer 254

Fast mercury determination in analytical applications

The LabAnalyzer is the perfect solution for quantitative determination of mercury in aqueous samples and sample digests in the laboratory.

 

In contrast to a typical multi-element AAS the LabAnalyzer 254 is specially designed for elemental mercury. This allows top performance in analytical applications. The use of a specially developed highly stable mercury lamp in connection with thermostat-controlled UV sensors results in a detection limit of a few ppt of mercury.

DOWNLOADS

FEATURES AND BENEFITS

  • Optical system does not require any readjustment
  • Short analysis time
  • Low reagent consumption
  • Automatic zero adjustment
  • Memory effects are minimised and a high sample throughput is possible thanks to specially selected materials for sample gas conducting components and heating of the optical bench.
  • For analyses according to DIN 38406-12 / EPA 7470A / EPA 7471A)
  • Measuring range:0.01 ppb … 10 ppb (10 ng/l to 10 µg/l)
  • Hg trap: Mercury vapors cannot escape into the lab
  • Upgradable to VM-3000 Mercury Vapor Monitor
  • Software for data acquisition, data visualisation and storage
  • Automatic Calibration Module
  • Special version with UT-3000 Ultratracer available

MAIN APPLICATIONS

  • Water samples: drinking water, waste water, ground water, surface water, sea water
  • Soil and sludge samples
  • Geological sample material
  • Waste samples: glass, construction rumble, contaminated liquids, wood
  • Incineration plant monitoring: smoke gas scrubber water, smoke gas analysis (e.g. to VDI 3868-2 VE)
  • Monitoring of foodstuffs
  • Clinical samples: urine, saliva
  • Chemical industry: environmental protection and quality control
  • Petrochemical industry
  • Scientific research

TECHNICAL INFORMATION

Measuring principle

Cold vapor atomic absorption (CVAAS) at 253.7 nm; Peak method

UV-Source

Electrodeless low pressure mercury lamp (EDL)

Stabilisation method

Reference beam method

Optical cell

Fused silica (Suprasil), l: approx. 230mm, heated approx. 45°

Sample volume

2 – 10 ml

Measuring ranges

liquid sample: 0,01 to 10 μg/l at 10 ml sample volume;

solid sample: 0,1 ng/g to 0,1 μg/g at 1 g sample weight

Sensitivity

approx. 5 ng/l resp. 0.05 ng absolutely

Reducing agent

Tin-II-chloride or sodiumborohydride

Signal outputs

analogue: 4-20 mA; serial: RS 232 / USB

Power supply

230 VAC / 50 Hz (110 VAC / 60 Hz option)

Power consumption

100 VA

Dimensions (W x H x D)

Photometer: 45 x 15 x 35 cm;

Reactor: 24 x 48 x 27 cm

Required floor space

approx. 70 x 50 cm (W x D)

Weight

approx. 10 kg